Hazardous-Area Approved DP1001 with Armor

Armored Intrinsically Safe 8 mm PRO Proximity Probes
DP100102
Standard Mount
A dimension drawing of a Standard Mount CTC PRO Line Hazardous-Area Approved DP1001 with Armor proximity probe.
DP100104
Standard Mount
A dimension drawing of a Standard Mount CTC PRO Line Hazardous-Area Approved DP1001 with Armor proximity probe.
Lengths AA, BB, and CC are configurable
Specifications
Temperature Range -49 to 350 °F (-45 to 177 °C)
Sensitivity -200 mV/mil (-7.87 V/mm) (1 mil = 0.001 in.)
Material FEP Cable Jacket with 316L Stainless Steel Probe Body
Case Thread 3/8-24 or M10x1
Wrench Flats 5/16 in. or 8 mm (Standard Mount), 7/16 in. or 10 mm (Reverse Mount)
Rating IP68
Connector SecureLock
Connector Pull Strength Up to 30 lbs.
Built to Order

Regulatory Approvals

US & Canada:
Class I Division 1 Groups A, B, C, D
Class II Division 1 Groups E, F, G
Class III
Class I Zone 0 AEx ia IIC
Class I Division 2 Groups A, B, C, D
Class II Division 2 Groups E, F, G
Ex ia IIC Ga
ATEX:
SIRA 16ATEX2084X
Ex ia IIC
International (IECEx):
IECEx CSA 16.0017X
Ex ia IIC
Temperature Codes:
T4 = -45°C to 100°C
T5 = -35°C to 85°C
IECEX Certified CSA Certified ATEX Certified CE Certified UKCA Certified ETL Certified
FOR REFERENCE ONLY, PLEASE REFER TO CONTROL DRAWING